vancomycin hydrochloride, USP ADD-VANTAGE Instructions For Use

()

INSTRUCTIONS FOR USE

To Use Vial in ADD-Vantage Flexible Diluent Container

To Open:

Peel overwrap at corner and remove solution container. Some opacity of the plastic due to moisture absorption during the sterilization process may be observed. This is normal and does not affect the solution quality or safety. The opacity will diminish gradually.

To Assemble Vial and Flexible Diluent Container:

(Use Aseptic Technique)

  1. Remove the protective covers from the top of the vial and the vial port on the diluent container as follows:
    1. To remove the breakaway vial cap, swing the pull ring over the top of the vial and pull down far enough to start the opening (See Figure 1.), then pull straight up to remove the cap. (See Figure 2.) NOTE: Do not access vial with syringe.
      Figure 1Figure 2
    2. To remove the vial port cover, grasp the tab on the pull ring, pull up to break the tie membrane, then pull back to remove the cover. (See Figure 3.)
  2. Screw the vial into the vial port until it will go no further. THE VIAL MUST BE SCREWED IN TIGHTLY TO ASSURE A SEAL. This occurs approximately 1/2 turn (180°) after the first audible click. (See Figure 4.) The clicking sound does not assure a seal; the vial must be turned as far as it will go. NOTE: Once vial is seated, do not attempt to remove. (See Figure 4.)
  3. Recheck the vial to assure that it is tight by trying to turn it further in the direction of assembly.
  4. Label appropriately.
    Figure 3Figure 4

To Reconstitute the Drug:

  1. Squeeze the bottom of the diluent container gently to inflate the portion of the container surrounding the end of the drug vial.
  2. With the other hand, push the drug vial down into the container telescoping the walls of the container. Grasp the inner cap of the vial through the walls of the container. (See Figure 5.)
  3. Pull the inner cap from the drug vial. (See Figure 6.) Verify that the rubber stopper has been pulled out, allowing the drug and diluent to mix.
  4. Mix container contents thoroughly and use within the specified time.
    Figure 5Figure 6

Preparation for Administration:

(Use Aseptic Technique)

  1. Confirm the activation and admixture of vial contents.
  2. Check for leaks by squeezing container firmly. If leaks are found, discard unit as sterility may be impaired.
  3. Close flow control clamp of administration set.
  4. Remove cover from outlet port at bottom of container.
  5. Insert piercing pin of administration set into port with a twisting motion until the pin is firmly seated. NOTE: See full directions on administration set carton.
  6. Lift the free end of the hanger loop on the bottom of the vial, breaking the two tie strings. Bend the loop outward to lock it in the upright position, then suspend container from hanger.
  7. Squeeze and release drip chamber to establish proper fluid level in chamber.
  8. Open flow control clamp and clear air from set. Close clamp.
  9. Attach set to venipuncture device. If device is not indwelling, prime and make venipuncture.
  10. Regulate rate of administration with flow control clamp.

WARNING: Do not use flexible containers in series connections.

Find vancomycin hydrochloride, USP ADD-VANTAGE medical information:

Find vancomycin hydrochloride, USP ADD-VANTAGE medical information:

Our scientific content is evidence-based, scientifically balanced and non-promotional. It undergoes rigorous internal medical review and is updated regularly to reflect new information.

vancomycin hydrochloride, USP ADD-VANTAGE Quick Finder

Prescribing Information
Download Prescribing Information

Health Professional Information

Instructions For Use

INSTRUCTIONS FOR USE

To Use Vial in ADD-Vantage Flexible Diluent Container

To Open:

Peel overwrap at corner and remove solution container. Some opacity of the plastic due to moisture absorption during the sterilization process may be observed. This is normal and does not affect the solution quality or safety. The opacity will diminish gradually.

To Assemble Vial and Flexible Diluent Container:

(Use Aseptic Technique)

  1. Remove the protective covers from the top of the vial and the vial port on the diluent container as follows:
    1. To remove the breakaway vial cap, swing the pull ring over the top of the vial and pull down far enough to start the opening (See Figure 1.), then pull straight up to remove the cap. (See Figure 2.) NOTE: Do not access vial with syringe.
      Figure 1Figure 2
    2. To remove the vial port cover, grasp the tab on the pull ring, pull up to break the tie membrane, then pull back to remove the cover. (See Figure 3.)
  2. Screw the vial into the vial port until it will go no further. THE VIAL MUST BE SCREWED IN TIGHTLY TO ASSURE A SEAL. This occurs approximately 1/2 turn (180°) after the first audible click. (See Figure 4.) The clicking sound does not assure a seal; the vial must be turned as far as it will go. NOTE: Once vial is seated, do not attempt to remove. (See Figure 4.)
  3. Recheck the vial to assure that it is tight by trying to turn it further in the direction of assembly.
  4. Label appropriately.
    Figure 3Figure 4

To Reconstitute the Drug:

  1. Squeeze the bottom of the diluent container gently to inflate the portion of the container surrounding the end of the drug vial.
  2. With the other hand, push the drug vial down into the container telescoping the walls of the container. Grasp the inner cap of the vial through the walls of the container. (See Figure 5.)
  3. Pull the inner cap from the drug vial. (See Figure 6.) Verify that the rubber stopper has been pulled out, allowing the drug and diluent to mix.
  4. Mix container contents thoroughly and use within the specified time.
    Figure 5Figure 6

Preparation for Administration:

(Use Aseptic Technique)

  1. Confirm the activation and admixture of vial contents.
  2. Check for leaks by squeezing container firmly. If leaks are found, discard unit as sterility may be impaired.
  3. Close flow control clamp of administration set.
  4. Remove cover from outlet port at bottom of container.
  5. Insert piercing pin of administration set into port with a twisting motion until the pin is firmly seated. NOTE: See full directions on administration set carton.
  6. Lift the free end of the hanger loop on the bottom of the vial, breaking the two tie strings. Bend the loop outward to lock it in the upright position, then suspend container from hanger.
  7. Squeeze and release drip chamber to establish proper fluid level in chamber.
  8. Open flow control clamp and clear air from set. Close clamp.
  9. Attach set to venipuncture device. If device is not indwelling, prime and make venipuncture.
  10. Regulate rate of administration with flow control clamp.

WARNING: Do not use flexible containers in series connections.

Medication Guide

Health Professional Information

{{section_name_patient}}

{{section_body_html_patient}}

Resources

Didn’t find what you were looking for? Contact us.

MI Digital Assistant

Chat online with Pfizer Medical Information regarding your inquiry on a Pfizer medicine.

Call 800-438-1985*

*Speak with a Pfizer Medical Information Professional regarding your medical inquiry. Available 9AM-5Pm ET Monday to Friday; excluding holidays.

Medical Inquiry

Submit a medical question for Pfizer prescription products.

Report Adverse Event

Pfizer Safety

To report an adverse event related to the Pfizer-BioNTech COVID-19 Vaccine, and you are not part of a clinical trial* for this product, click the link below to submit your information:

Pfizer Safety Reporting Site

*If you are involved in a clinical trial for this product, adverse events should be reported to your coordinating study site.

If you cannot use the above website, or would like to report an adverse event related to a different Pfizer product, please call Pfizer Safety at (800) 438-1985.

FDA Medwatch

You may also contact the U.S. Food and Drug Administration (FDA) directly to report adverse events or product quality concerns either online at www.fda.gov/medwatch or call (800) 822-7967.